• UL 61010-2-091

UL 61010-2-091

Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-091: Particular requirements for cabinet X-Ray systems

Underwriters Laboratories , 08/08/2014

Publisher: UL

File Format: PDF

$131.00$263.21


Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 2-091: Particular Requirements for Cabinet X-Ray Systems
UL 61010-2-091 (IEC 61010-2-091:2012)
1 Scope and object
This clause of Part 1 is applicable, except as follows:
1.1.1 Equipment included in scope
Replacement:Replace the text with the following:A CABINET X-RAY SYSTEM is a system that contains an X-ray tube installed in a cabinet which, independently of existing architectural structures except the floor on which it may be placed, is intended to contain at least that portion of a material being irradiated, provide radiation attenuation, and exclude personnel from the interior during generation of X-radiation.These CABINET X-RAY SYSTEMS are used in industrial, commercial, and public environments, for example, to inspect materials, to analyze materials, and to screen baggage.

1.1.2 Equipment excluded from scope
Addition:Add the following new items to the list:
aa) equipment intended to apply X-radiation to humans or animals;
bb) equipment incorporating an X-ray tube but not incorporating complete shielding against X-radiation hazards, such as:
  • equipment intended to be used within a shielded room which excludes personnel during operation;
  • equipment intended to be used with separate portable or temporary shielding;
  • equipment intended to produce an emerging beam of X-radiation.
1.2.1 Aspects included in scope
Addition:Add the following text at the end of the first paragraph:This part of IEC 61010 specifies requirements for the design and methods of construction of CABINET X-RAY SYSTEMS to provide adequate protection for OPERATORS, bystanders, trained service personnel, and the surrounding area against unintentionally-emitted X-radiation and from mechanical HAZARDS related to their conveyors.

UL 61010-2-091 History

UL 61010-2-091

UL 61010-2-091

$324.00 $649.00

UL 61010-2-091

UL 61010-2-091

$131.00 $263.21

More UL Standards PDF

TIA TSB-88.3-E

TIA TSB-88.3-E

$106.00 $212.00

UL 248-16

UL 248-16

$79.00 $158.20

TIA TSB-102.CBBK-B

TIA TSB-102.CBBK-B

$38.00 $77.00

UL 61215-1-4

UL 61215-1-4

$89.00 $178.92